LAND CT3002A(5V5mA&50mA8C1U)Specifications
| CT3002A 5V5mA&50mA8C1U | |||
| Items | Indicators | ||
| Input Power Source | AC 220V±10%/50HZ (AC 110V customizable) | ||
| Max. Output Power | 30W | ||
| Resolution | AD: 16bit; DA: 12bit | ||
| Input Impedence | ≥1GΩ | ||
| Voltage | Test Range | Charge: 0V-5V; Discharge: -5V-5V | |
| Min. discharge voltage | 0V | ||
| Measurement Accuracy | 0.05%FS | ||
| Current | Test Range | 10uA-50mA | |
| Error | ≤2.5uA | ||
| Measurement Accuracy | 0.05%FS | ||
| Output Power of Each Channel | 0.25W | ||
| Time | Current Response Time | ≤10mS | |
| Step Interval | 50mS | ||
| Available Setup | ≤365*24 Hours/Step | Time Format 00:00:00
(h, min, s) |
|
| Data Record | Conditions | △t: 20mS-7200S | |
| △U: 2mV-5V | |||
| △I: 2uA-2000A | |||
| Frequency | 10Hz | ||
| Charge | Model | HOLD, CC, CV, CP, DCIR, CRATE | |
| Constraints | Voltage, Current, Time, Capacitance etc almost 20 kinds | ||
| Discharge | Model | HOLD, DC, DV, DP, DR, DRATE | |
| Constraints | Voltage, Current, Time, Capacitance etc almost 20 kinds | ||
| Cycle | Times | 1-65000 | |
| Max. Steps | 1100 | ||
| Max.Nesting Layers | 3 | ||
| Protection Mechanism | Power outage, Anti-reverse connection | ||
| Settable protective mode: Upper/Lower limit of voltage, Upper/Lower limit of current, Anti-overcharge, Anti-overdischarge, CC Anti-voltage change, CV Anti-current change, Correlate channel temperature/voltage/pressure | |||
| Testing Mode | Four Electrodes (Also can be applied in tri-electrode) | ||
| Channels | Independent | ||
| Noise | ≤75dB | ||
| Communication Mode | RS422 | ||
| Data Format | CSV, EXCEL, TXT | ||
| Environment | Temperature: -10℃-50℃ (Best 23±2℃)
Humidity: ≤95% |
||